190
Chapter 4
Calibration
Changing the Calibration Kit Definition
Step 3.
Select
Sub Class3
.
Step 4.
Press
TRL Line/Match - Set All - LINE >7G - Return
.
Step 5.
Press
and check that the name you specified is selected as Cal Kit.
By assigning Match and Line 1/2 to subclass 1/2/3 respectively, you can calibrate 3
standards with different frequency bands for TRL line/match calibration.
Setting options for TRL calibration
This section describes how to set the reference impedance and measurement data of the
standard used for calculating the calibration plane for TRL calibration.
Step 1.
Press
.
Step 2.
Press
Cal Kit
, and select a calibration kit.
Step 3.
Press
Modify Cal Kit
.
Step 4.
Press
TRL Option
.
Step 5.
Press
Impedance
to select the reference impedance.
Softkey
Function
Line
Calculates the calibration coefficient using the characteristic impedance of the
line standard as the reference impendence. The impedance of the line standard
is used as the center of the Smith chart, and the calibration coefficient is
calculated using the line standard’s as the direction.
System
Calculates the calibration coefficient using the system impedance as the
reference impendence. Select this when the line impedance and the test port
impedance differ.
Step 6.
Press
Reference Plane
to select the measurement data of the standard used for calculating
the reference plane.
Softkey
Function
Thru
Uses the length of the thru/line standard to calculate the calibration plane.
Reflect
Uses the reflection coefficient of the reflection standard to calculate the
calibration plane.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...