Table 3‑2. Electrical specifications (Continued)
Description
Specification
Setpoint change response
time
Analog
User
interface
Analog response time is measured from step setpoint change to
90% change of the RF output voltage.
• RF rise time < 30 ms over full power range into a broadband
load
• RF fall time < 30 ms over full power range into a broadband
load
Setpoint change response
time
Serial
Host
interface
Serial interface response time is measured upon receipt of an AE
Bus command to 90% change of the output RF voltage.
• RF rise time < 25 ms over full power range into a broadband
load
• RF fall time < 25 ms over full power range into a broadband
load
Spurious signals in full
frequency and power
range on 50 Ω load
- 50 dBc
Harmonics in full
frequency range at
maximum power on 50 Ω
load
- 30 dBc
Maximum allowable
forward power incident
upon the RF outputs from
external sources/ load
50 W minimum between 2 MHz and 300 MHz
Zero/Low setpoint
RF output is disabled if setpoint in power control mode is less than
5 W. RF output is disabled if setpoint in external voltage control
mode is less than 20 V.
RF ON
LED on the front panel remains on.
RF ramp rate control
The rise and fall rate of the RF output can be programmed via the
serial
Host
interface. This function is enabled for RF ON and
setpoint up or down changes via the serial
Host
interface.
For safety reasons, an RF OFF command will not ramp down the
output. The rate can be defined in watts per second or absolute time
through the serial
Host
interface.
Watts per second mode can be set in 1 W increments.
Absolute time mode range in 1 ms increments:
• Minimum ramp time: 1 ms
• Maximum ramp time: 30000 ms
The ramp rate parameters are not allowed to change while a setpoint
ramp is currently in progress.
Advanced Energy
®
Paramount
®
MF 2 kW Generator
57020114-00E
Specifications
3‑6
Содержание Paramount MF 2 kW
Страница 2: ...Paramount MF 2 kW Generator User Manual January 2018 57020114 00E...
Страница 31: ...Figure 3 2 Pressure drop Advanced Energy Paramount MF 2 kW Generator 57020114 00E Specifications 3 12...
Страница 103: ...3710 Figure 5 5 Left side Advanced Energy Paramount MF 2 kW Generator 57020114 00E Installation Setup and Operation 5 6...
Страница 105: ...3712 Figure 5 7 Top side Advanced Energy Paramount MF 2 kW Generator 57020114 00E Installation Setup and Operation 5 8...