suitable chassis ground connection made to this stud prevents or minimizes
radio frequency interference.
Connecting Cooling Water
This generator is water cooled. Do not operate the unit until you have connected
cooling water and met the cooling requirements.
CAUTION:
If you connect the cooling water on multiple units in series, be sure that input
water temperature to all units is less than the maximum input water
temperature.
CAUTION:
If you connect the cooling water on multiple units in a series, be sure that the
water flow rate to all units meets the unit specifications.
CAUTION:
Do not use deionized water for cooling purposes. Deionized water causes
both corrosion and erosion of cooling manifolds.
CAUTION:
Do not apply more than 54.2 Nm (40 ft-lb) to water fittings. Water leakage
and damage to the unit could result.
The water fittings used on the Paramount MF 2 kW generator are SS female 3/8″
Swagelok.
TO CONNECT COOLING WATER
1. Make input and output water connections and tighten securely.
2. Turn on the water and ensure that there are no leaks.
3. Be sure that the flow rate, pressure, and temperature are within the minimum
specifications required to operate the generator.
Advanced Energy
®
Paramount
®
MF 2 kW Generator
57020114-00E
Installation, Setup, and Operation
5‑11
Содержание Paramount MF 2 kW
Страница 2: ...Paramount MF 2 kW Generator User Manual January 2018 57020114 00E...
Страница 31: ...Figure 3 2 Pressure drop Advanced Energy Paramount MF 2 kW Generator 57020114 00E Specifications 3 12...
Страница 103: ...3710 Figure 5 5 Left side Advanced Energy Paramount MF 2 kW Generator 57020114 00E Installation Setup and Operation 5 6...
Страница 105: ...3712 Figure 5 7 Top side Advanced Energy Paramount MF 2 kW Generator 57020114 00E Installation Setup and Operation 5 8...