3.39 Bowl IR LED Assembly Removal
1)
Remove the front and rear covers (3.2, 3.3).
2)
Remove the tape holding the IR LED assembly wires to the outside of the
bowl, and unplug the connector.
3)
For the right-hand bowl IR LED, slide the connector and wires out towards
the front, between the bowl and the baffle and the bottom of the bowl.
Replacement Notes:
P
Tighten the left/right IR LED mounting screws only finger tight until after
adjustment has been performed.
Follow-up Checks/Adjustments: