Humphrey Field Analyzer II - i
Parts Removal/Replacement
Confidential and Proprietary
52235C1105
3 - 45
Field Service Guide
3)
Disconnect the "X-MOTOR" cable (P4) from the Patient Support Interconnect
PCB. Cut any cable ties holding this cable to the Patient Support Interconnect
PCB.
4)
Rotate the motor shaft to position the lead screw nut close to the X-motor
(Figure 3-10).
5)
Loosen the setscrews holding the lead screw to the motor shaft.
6)
Grasp the chinrest and move it away from the X-motor until the lead screw
disengages from the motor shaft.
7)
Remove the four screws securing the X-motor to the bracket, and remove the
motor.
FIGURE 3-10. Disconnecting the X-Motor Lead Screw from the Motor Shaft
Replacement Notes:
P
Be sure to reinstall a nylon washer on the motor shaft before attaching the lead
screw. The lead screw should be seated on the motor shaft all the way up to the
motor body before tightening the setscrews.
P
Before installing the belt:
After the lead screw is secured to the motor shaft, rotate the shaft to position the
lead screw nut up against its limit nearest to the motor. Then set the headrest
horizontal lead screw nut to the same relative position (check for the same amount
of thread showing on both lead screws).
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...