FA2i-023
Page 3
Repair of Corrupted Calibration Data disk
1. Boot up instrument
2. Insert Calibration Backup Disk into the instrument floppy drive.
3. From the HFA2i instrument
Diagnostic
menu select
SAVE
to save Calibration Data onto the
original Calibration Backup Disk.
4. Take out the disk and follow instructions 2-14 of the Verification of Calibration Backup Disk
procedure above.
5. Initial and date the disk label.
6. Return the Calibration Backup Data disk to the holder in the cable compartment of the instrument.
7. Turn off instrument and reboot.
8. Verify the configuration settings are correct.
Manufacturing Verification of Calibration Backup Disk File Implementation
Instruments that have been manufactured with the new Calibration Backup Disk verification process start
with the following serial numbers:
720i-5838
740i-11446
745i-4559
750i-9393
Correction of Writing Corrupted Backup Data Disks
The next HFA2i software release will have provisions that will reduce or eliminate the possibility of writing
corrupted data to the Calibration Backup disk. This bulletin will be revised at that time.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...