Humphrey Field Analyzer II - i
Appendix C
Confidential and Proprietary
52235C1105
C - 9
Field Service Guide
On the Transmitting
HFA II - i
25)
To proceed, select YES. After all tests are transmitted, a pop up dialog box will appear
advising you of the number of tests that were successfully transferred.
26)
Press the Main Menu icon to exit this function.
C.6
HFA II - i
To Ensemble© Serial Data Transfer
Data will normally be transmitted from the
HFA II - i
to an Ensemble PC via the network
interface. As a temporary means of data transfer or as a troubleshooting technique, serial data
transfer from the
HFA II - i
to Ensemble is possible. Data can be transferred in both the Serial I
transfer mode and the Serial II transfer mode. You must have the
HFA II - i
in the Serial II
transfer mode to ensure that all of the test information is passed to the Ensemble program.
Transmitting
HFA II - i
Tests to Ensemble
1)
Ensure that the Serial II Transfer Disk is loaded. Check the destination window of the
Transfer Files selection on the Files Functions Menu. If not, load the Serial II Transfer
Disk as described in Appendix section C.2.
2)
Ensure that the data transfer cable is connected between the
HFA II - i
DATA
TRANSFER port and the appropriate serial port of the Ensemble computer. See the
Ensemble documentation for details.
3)
Ensure that Ensemble is prepared to accept data from the
HFA II - i
. See the Ensemble
documentation for details.
4)
From the Main Menu, select
System Setup
.
5)
Select the
Save/Transmit
option.
6)
Select
RS-232 Options
.
7)
On the RS-232 Options screen, select the following settings:
A) Set Baud Rate to 9600
B) Parity to Even
C) Data Bits to 7
D) Stop Bits to 1
8)
Press
Proceed
. The Transmit/Save screen will then appear.
9)
Press the
Main Menu
icon in the upper right corner of the screen.
10) From the
Main Menu
, select
File Functions
.
11)
Select
Transfer Tests
. A dialog box titled “Disk Options” will appear on screen.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...