Humphrey Field Analyzer II - i
PM and System Checkout
Confidential and Proprietary
52235C1105
2 - 7
Field Service Guide
19)
Check the operation of the rear fan, and clean the filter
20)
Check that the date is correct and the time is within 2 minutes.
21)
Select a patient test. Change parameters. Verify that the central fixation LED,
inner diamond, and outer diamond all function.
22)
Turn foveal threshold on (except Model
720i
). Select size III. Start the test. Verify
that the size III spot is in the center of the inner fixation diamond. Ensure that the
spots are sharply focused and no halo is visible. Test the patient button for
operation.
23)
Recall a patient test from your test floppy disk. Display and print the test.
24)
Recall a test from the hard disk. Display and print the test.
25)
Duplicate a patient test from your test floppy disk to a formatted floppy. Display
the copied test.
26)
Copy a patient test from your test floppy disk to the hard drive. Delete that test
from the hard disk.
27)
Perform a backup using the Magneto-Optical drive to verify the backup/restore
procedure.
28)
Turn power off/on and wait for the Main Menu. No errors should be reported.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...