![background image](http://html1.mh-extra.com/html/zeiss/humphrey-hfa-ii-i-series/humphrey-hfa-ii-i-series_field-service-manual_3332218147.webp)
Humphrey Field Analyzer II - i
Adjustment / Calibration
Confidential and Proprietary
52235C1105
4 - 45
Field Service Guide
4.9.7
Camera Focus
Purpose
This procedure adjusts the camera lens for optimum focus. It is not normally necessary
to refocus the camera unless you have done replacements that affect the camera lens
position (i.e., bowl replacement, fixation LED replacement, camera replacement).
Special Tools/Equipment Required
P
Tool Stand
P
Fake Eye with Camera Focus Target attached
P
RTV
Prerequisites
Before starting Camera Focus, the following must be done:
1)
Check/adjust the Trial Lens Holder (Section 4.9.6)
2)
Check/adjust the Camera Position/Size (Sections 4.8.6.1 and 4.8.6.2).
Procedure
1)
Install the tool stand on the instrument, with fake eye attached and height adaptor
removed. Push the tool stand forward slightly on the chinrest so that the metal
tips on the Tool Stand Alignment Adaptor make contact with the headrest (Figure
4-2). This is the position the tool stand should be in when performing the
procedure.
2)
From the Calibration main menu, select
CAMERA
, then
POSITION
.
3)
Using the chinrest control switch, align the camera focus target in the video insert.
4)
Gently rotate the camera lens manually and observe the focus target for best
resolution obtainable. (You may need to use two fingers on opposite sides of the
lens in order to rotate the lens without binding.)
5)
Apply RTV at the point indicated in Figure 4-18 to secure the lens in position.
NOTE:
Squeeze the RTV onto a piece of waste paper and allow it to thicken slightly
and become tacky before you apply it to the lens.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...