Parts Removal/Replacement
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
3 - 16
52235C1105
6)
Pull the drive housing assembly straight out until the cable connections at the back are
accessible.
7)
Disconnect the cables and power connectors from the drives.
FIGURE 3-3. Unlatching the Drive Housing Bezel
Replacement Notes:
P
Before installing a new drive, verify that it is strapped correctly (Refer to Field Service
Bulletin FA2i-003(x)).
P
Reconnect the cables to the drives before reinstalling the drive housing assembly.
P
Replacement hard disks are preformatted. You will need to load the application
software, and restore configuration and calibration data. If the calibration backup
diskette is not available, a complete system calibration is required. Patient data will
need to be restored, preferably by the customer.
P
If it becomes necessary to reformat a hard drive, refer to Appendix L for instructions.
Follow-up Checks/Adjustments:
See Table 3-1.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...