Adjustment / Calibration
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
4 - 38
52235C1105
Calibration Procedure
NOTE:
Use the pencil eraser tip, or similar soft-ended implement, during this procedure.
1)
From the additional setup menu, select
Touch Screen Calibration
. If you are
unable to access the additional setup menu, you can power-up the instrument while
depressing the patient button until the touch screen calibration menu comes up. This
will take approximately 2½ minutes.
2)
Follow the instructions displayed on the screen.
When you are instructed to press the rectangle in the Top Left Corner, note exactly
where it is within the rectangle that you touch the screen. This will enable you to
"follow the error," (as described above) if another attempt at calibration is necessary.
NOTE:
press the rectangle only
once
; then
wait
for the screen to display the instruction to
press the rectangle in the Bottom Left Corner.
When you touch the rectangle, again note exactly where you touch the screen.
4.9.5
CRT Adjustments
Special Tools / Equipment Required
P
Plastic Slot-Tip Alignment Tool
P
Operator Panel Extension/Support Tool
P
Extension Cables
P
CRT Overlay Template (refer to service bulletin HFA2-029)
WARNING:
During these adjustments you will be working closely
around the back of the CRT with power applied. Work carefully to
avoid the areas where the CRT high voltage is present.
Procedures
NOTE:
Because some of the CRT adjustments interact, they should be performed in the
order presented here.
All of the CRT adjustment control pots are located on the CRT Driver PCB (Figure
4-13). The name of each pot is labeled on the PCB.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...