5)
Select
Save
to store the serial number and return to the Configuration Menu.
(Selecting Quit will not store the serial number and will return to the Configuration
Menu.)
6)
Switch OFF the instrument and remove the Service Key.
7)
Switch ON the instrument and verify that the correct serial number appears on the
screen during the boot-up process.
G.3 Setting the Model/Hardware Options
NOTE:
The hardware options codes listed in this section are for C1.1 system software.
Changes in future revisions of software may result in changes to the hardware options.
This procedure is designed to set up the basic hardware configuration by setting the model
number of the instrument. This is required whenever the CAL/Config Data has just been
initialized.
For Model 720i
1)
Set up the DST for use as described in Appendix A.2, under Diagnostic Support Tool.
2)
From the DST Main Menu select
Set Model/Hardware Options
. The
Model/Hardware Options Menu appears.
3)
Select
Set Model 720
i
Defaults
, a test log menu appears and displays the
model 720
i
default settings.
4)
Select
Continue
. The Model
720i
hardware Options are set and the Model/Hardware
Options Menu is displayed.
5)
Switch OFF the instrument and remove the Service Key.
6)
Switch ON the instrument. Select the “i” screen and verify that the correct model
number appears on the boot-up screen.
7)
Proceed to “Setting the Software Options.”
For Model 740
i
1)
Set up the DST for use as described in Appendix A.2, under Diagnostic Support Tool.
2)
From the DST Main Menu select
Set Model/Hardware Options
. The
Model/Hardware Options Menu appears.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...