SERVICE
BULLETIN
Instrument:
Humphrey Field Analyzer II - i
Subject:
Differences between HFA II and HFA II - i
No.:
FA2i-002
Date:
15 Jul 01
Status:
Informational
Pages:
1
This service bulletin will summarize some of the major differences between the HFA II system and the
HFA II - i system. If you are trained on the HFA II system, the similarities between the two models are
close enough that additional classroom training is not required.
Computer
HFA II
HFA II - i
CPU
Motorola 68020 33 MHz
Intel Celeron 433 MHz
Hard Drive Controller
SCSI
IDE
Memory
4MB
64MB
Communications
3 Serial Ports
1 Serial Port, Network Port,
2 USB Ports (Future Use)
The CPU plugs into the Backplane PCB and is enclosed in an ESD enclosure. The Backplane PCB
controls the Patient Switch, Touch Screen, Camera, Frame Grabber and Video.
Drive Module
HFA II
HFA II - i
Floppy
TEAC
Same drive with different jumper settings
Hard Drive
SCSI
IDE
Data Backup
Tape Drive
Magnetic Optical Drive
A new power supply is used, which has additional vcc power capacity. The HFA II and HFA II - i power
supplies are not interchangeable. The HFA II - i uses a new Motor Driver Board, which is backward
compatible to the HFA II. The back cover of the instrument has been re-designed to fit over the CPU
enclosure and provides access to peripheral connectors.
Service Changes
All calibration procedures for the HFA II - i are identical to the HFA II. The HFA II - i saves all calibrations
to the Hard Drive. The calibration file is also saved to a floppy disk and can be restored through the
calibration menu. The back cover of the instrument provides a special storage slot for the calibration
backup diskette.
The HFA II - i utilizes a new DST tool for changing the configuration of the instrument. Additionally, you
may need to change BIOS settings for future upgrades to the system. The CPU diagnostics are no longer
available; the Quick Tech 2000 Pro Tool will be used for HFA II - i CPU diagnostics.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...