SERVICE
BULLETIN
Instrument:
Humphrey Field Analyzer II-i
No.:
FA2i-022
Subject:
Start up Problems
Date:
24 Jun 03
Status:
Next Visit
Pages:
1
Problem
HFA2i instruments with the new CPU (P/N 55392), using application revision 3.4 software, may fail to
reach the main menu on startup. The display will be blank and the instrument will give no indication that a
failure has occurred.
Cause
Failures in the initialization of the RAM disk are one possible source for the blank display on startup.
Solution
Changes have been made to the application software in release 3.4.5. Load the application software
revision 3.4.5 to install the new RAM disk initialization routine and eliminate the start up problem. If the
software upgrade does not correct the start up problem, use the techniques found in the HFA2i service
manual to troubleshoot and repair the unit.
Instruments that have been manufactured with the new CPU and software revision 3.4, and require the
3.4.5 software, are in the following serial number ranges:
720i-5815 through 720i-5838
740i-11330 through 740i-11446
745i-4546 through 745i-4559
750i-9376 through 750i-9393
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...