Humphrey Field Analyzer II - i
Appendix H
Confidential and Proprietary
52235C1105
H - 1
Field Service Guide
Appendix H. Calibration Printouts
Contents:
C
Cal / Wedge Printout
C
Automated Light Intensity Verification Printouts
C
White Light Verification Data
C
Blue-Yellow Verification Data
H.1 Cal / Wedge Printout
The numbered definitions below (1, 2, 3, etc.) are keyed to the numbered elements of the
sample calibration printout shown in Figure H-1.
1 = Date/time of calibration printout, instrument serial number, and hardware option
number.
2 = NVM Data
The numbers that appear in the printout represent the current calibration values stored
in Cal/Config Data for each item listed, for this particular instrument. A brief definition
of each item is given below, followed by a table that lists the acceptable tolerance range
for each item.
Proj . . . . . . .
represents voltage
*
(left column) and intensity
*
(right column) of the
projection lamp when a white stimulus is projected.
Bowl
. . . . . . represents voltage
*
and intensity
*
for the white bowl lamps.
Blue
. . . . . . . represents voltage
*
and intensity
*
of the projection lamp when a blue
stimulus is projected.
Yellow
. . . . . represents voltage
*
and intensity
*
for the yellow bowl lamp
Right Mode
. microsteps of positioning
Left Mode
. . microsteps of positioning
Offset Fix
. . . microsteps of positioning
Focus
. . . . . . microsteps of positioning
Shutter
. . . . microsteps of positioning
Aperture
. . . microsteps of positioning for each of the 5 positions of the aperture
wheel
Color
. . . . . . microsteps of positioning for each of the 5 positions of the color wheel
*
Voltage and intensity are expressed here as relative numbers, not actual units of voltage or
intensity (such as volts or foot-lamberts). Typically, any high round number (e.g., 30000) that
appears here in the printout is the default value for a parameter that has not been calibrated.
It is possible but highly unlikely that a calibrated value would fall precisely at a high round
number.
3 =
Voltage drop across projector lamp when white stimulus is projected, and when blue
stimulus is projected.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...