Humphrey Field Analyzer II - i
PM and System Checkout
Confidential and Proprietary
52235C1105
2 - 5
Field Service Guide
HFA II - i
SYSTEM CHECKOUT CHECKLIST
Your Name: ____________________
HFA II - i
S/N: ____________________
Date: ____________________
NOTE:
Following any calibration procedure that has calibration values stored in the
Cal/Config Data, storage of the new values must be verified by cycling the power off-on
before you perform final system checkout
. If values obtained by the HFA II - i during
calibration are not within an expected range, those values are held only in temporary
storage and not written to the Cal/Config Data. When power is turned off, these values
are erased and calibration reverts to its old status.
The
HFA II - i
System Checkout Checklist Guidelines —
#
Part I —
The items described in Part I must be completed for
All
service calls.
#
Part II —
The items described in Part II must be completed when any of the following
service conditions apply:
-- Performing an instrument PM
-- Recalibrating either the White/White or Blue/Yellow light intensities
-- Replacing/initializing the Cal/Config Data
-- Cleaning/replacing the brightness detector, ND wedges, or the color wheel
-- Replacing the motor driver PCB
This checklist assumes that all required service has been performed on the
HFA II - i
. Place a
checkmark next to each item if okay. If an item does not apply to the
HFA II - i
under service,
indicate N/A (not applicable).
Part I -- For All Service Calls
(Refer to the above System Checkout Checklist Guidelines.)
1)
If a repair has been performed, verify that all required adjustments/checks have
been performed per Table 3-1.
2)
Acquire and print the results of your foveal test. Check that the foveal value
obtained is within ±2 dB of your known foveal value. (4.8.1.1)
3)
Print the Cal/Wedge printout. Verify that all values are within specification.
(Appendix H)
4)
Print the Unit Configuration printout. Verify that all entries are accurate, based on
instrument serial number and model. (Appendix J).
5)
Print the System Log printout and then clear the log. Verify that all reported error
conditions have been evaluated/corrected. (Appendix J)
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...