![background image](http://html1.mh-extra.com/html/zeiss/humphrey-hfa-ii-i-series/humphrey-hfa-ii-i-series_field-service-manual_3332218429.webp)
Page 4
FA2i-015
Instructions on how to evaluate test dates
Main Menu>System Setup>Additional Setup>Diagnostics>Enter password 1234>Evaluate Test
Dates> Select the Source (Hard Drive or Floppy). After reviewing the confirming message, if you
still wish to perform the evaluation, select “Yes”. If you do not want to perform the evaluation at this
time, select “Cancel”.
Note: You may wish to reprint the visual field test results with the correct test date recorded. Please
be aware that merely installing the new system software revision 3.2 will
not
correct the files of
previously conducted tests. You must follow the steps as indicated in the “Action Required” section
above
to correct the files of previously conducted tests.
Note: New Feature in Software Revision 3.2 - Internal Self-Test for Date Checking
System software revision 3.2 includes an automatic self-test for date validation. Although the new
self-test is set up to run unnoticed in the background, you may see a pop-up window that displays the
message: “Please verify that today’s date is shown correctly below” along with the date the
instrument is set to. If you receive this message and the date displayed is correct, press “Yes”. If the
date is incorrect, press “No”; input the appropriate date at the pop up menu, then select “Enter”.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...