52235C1105
POWER ENTRY AND POWER SUPPLY
Note
The switching power supply accepts AC power input voltages of 90 to 132/180 to
264 V AC. The fuse requirements are the same for all input voltages.
Notes:
1.
Layout is shown as viewed from left side of instrument.
2.
The HFA II Power Supply cannot be used in the HFA II-i.
DC VOLTAGE MEASUREMENTS
Voltage
Tolerances
+5 V
4.75 – 5.25
±12 V
±11.4 – 12.6
+24 V
22.8 – 25.2
FIGURE 6-8. Power Entry and Power Supply