![background image](http://html1.mh-extra.com/html/zeiss/humphrey-hfa-ii-i-series/humphrey-hfa-ii-i-series_field-service-manual_3332218268.webp)
Appendix C
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
C - 8
52235C1105
• If you chose
Name
:
To select all files on the drive, press
Enter
, and
Select All
. All files on the
disk will be marked with a “
9
” to indicate that they have been selected.
• If you search files by patient name:
Enter the name - or just the first few characters - of the patient’s last name whose test(s)
you wish to transfer. Select the test(s) you wish to transfer. All files selected will be
marked with a “
9
” to indicate that they have been selected.
• If you chose Date:
The file directory will be displayed in reverse chronological order. Enter a specific date
(mm-dd-yyyy), then choose
Enter.
The file directory will display tests taken on the
date you selected. Select the test(s) you wish to transfer. All files selected will be marked
with a “
9
” to indicate that they have been selected.
14)
Press
Proceed
. A pop up dialog box will appear advising you of the number of tests
selected to be transferred.
STOP HERE!
Do Not choose
Yes
to start the transfer
process.
On the Receiving
HFA II - i
:
15)
From the Main Menu, select System Setup.
16)
Select the Save/Transmit option.
17)
Select RS-232 Options.
18)
On the RS-232 Options screen, select the following settings:
A) Set Baud Rate to 9600
B) Parity to Even
C) Data Bits to 7
D) Stop Bits to 1
19)
Press Proceed. The Transmit/Save screen will then appear.
20)
Press the Main Menu icon in the upper right corner of the screen.
21)
From the main menu, select File Functions
22)
Select Transfer.
23)
Select HFA II Serial Cable as the Source.
24)
Choose Hard or Floppy as the Destination, then Proceed.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...