Parts Removal/Replacement
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
3 - 28
52235C1105
P
When re-inserting the Central Fixation LED, you must line up the notch on the
LED with the flat portion of the holder. You will hear a click when the LED moves
into position.
P
Refer also to the camera replacement notes in Section 3.12.
Follow-up Checks/Adjustments:
See Table 3-1.
3.17 Projection Assembly Removal
Caution —
The aperture wheel and ND wedges on the projection
assembly are delicate and easily damaged —
work carefully!
Any
fingerprints or contamination on the optical area of the film wedge may
necessitate replacement of the wedge; it cannot be cleaned, other than to
blow off any dust.
1)
Remove the front and rear covers (3.2, 3.3)
2)
Loosen the screw holding the finger guard/baffle plate just behind the film wedge
(Figure 3-5). Easiest access to the screw is through the hole in the film wedge.
FIGURE 3-5. Removing the Finger Guard/Baffle Plate