Troubleshooting
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
5 - 26
52235C1105
ERROR MESSAGE
PROBLEM
SOLUTION
Printer is offline or is
disconnected.
The HFA II was not able to
communicate with the printer.
Ensure that the Printrex
printer “green” power light is
on.
Check the printer power
connection and the data
cable. Reseat the both cables.
Check for the proper printer
selection in the System Set Up
menu.
Remove the paper carrier.
Blow any paper debris from
the IR LED hole in the lower
right corner of the Printrex
printer door opening.
Reseat the paper carrier and
paper as far to the right as
possible.
Perform the Printrex self test
described in section 5.6 of this
manual.
As a final repair action, try
another printer, then CPU
PCB.