Appendix C
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
C - 6
52235C1105
• If you search files by patient name:
Enter the name - or just the first few characters - of the patient’s last name whose test(s)
you wish to transfer. Select the test(s) you wish to transfer. All files selected will be
marked with a “
9
” to indicate that they have been selected.
• If you chose Date:
The file directory will be displayed in reverse chronological order. Enter a specific date
(mm-dd-yyyy), then choose
Enter.
The file
directory will display tests taken on the
date you selected. Select the test(s) you wish to transfer. All files selected will be marked
with a “
9
” to indicate that they have been selected.
14)
Press
Proceed
. A pop up dialog box will appear advising you of the number of tests
selected to be transferred.
STOP HERE!
Do Not choose
Yes
to start the transfer
process.
On the
HFA II - i
:
15)
From the Main Menu, select
System Setup
.
16)
Select the
Save/Transmit
option.
17)
Select
Serial Receive
option.
18)
On the
RS Receive Setup
options screen, select the following settings:
A) Set Baud Rate to 9600
B) Parity to Even
C) Data Bits to 7
D) Stop Bits to 1
19)
Press
Proceed
. The Transmit/Save screen will then appear. Press
Proceed
.
20)
The
System Setup
menu will appear.
21)
From the
System Setup
, select
File Functions
.
22)
Select
Transfer Tests
.
23)
Select
HFA Serial Cable
as the Source.
24)
Choose
Hard
or
Floppy
as the Destination, then
Proceed
.
On the HFA II
25)
To proceed, select
YES
.
26)
After all tests are transmitted, a pop up dialog box will appear advising you of the
number of tests that were successfully transferred.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...