Humphrey Field Analyzer II - i
Appendix G
Confidential and Proprietary
52235C1105
G - 5
Field Service Guide
3)
Select
Set Model 740
i
Defaults
, a test log menu appears and displays the
model 740
i
default settings.
4)
Select
Continue
. The Model 740
i
hardware Options are set and the Model/Hardware
Options Menu is displayed.
5)
Switch OFF the instrument and remove the Service Key.
6)
Switch ON the instrument. Select the “i” screen and verify that the correct model
number appears on the boot-up screen.
7)
Proceed to “Setting the Software Options.”
For Model 750
i
1)
Set up the DST for use as described in Appendix A.2, under Diagnostic Support Tool.
2)
From the DST Main Menu select
Set Model/Hardware Options
. The
Model/Hardware Options Menu appears.
3)
Select
Set Model 750
i
Defaults
, a test log menu appears and displays the
model 750
i
default settings.
4)
Select
Continue
. The Model 750
i
hardware Options are set and the Model/Hardware
Options Menu is displayed.
5)
Switch OFF the instrument and remove the Service Key.
6)
Switch ON the instrument. Select the “i” screen and verify that the correct model
number appears on the boot-up screen.
7)
Proceed to “Setting the Software Options.”
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...