![background image](http://html1.mh-extra.com/html/zeiss/humphrey-hfa-ii-i-series/humphrey-hfa-ii-i-series_field-service-manual_3332218288.webp)
Appendix G
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
G - 6
52235C1105
G.4 Setting the Software Options
NOTE:
Setting the software options is required whenever the hardware option number has
been changed or the Cal/Config Data has just been initialized. Please note that the DST
cannot be used to set the software option number unless the instrument is being repaired at
the Carl Zeiss Meditec factory. All other repair situations must follow the procedure
described below.
The software option number is automatically created during the installation of application
software. The software option number is based on the hardware option number already
written into the Cal/Config Data. Ensure that the correct hardware option number has been
written to the Cal/Config Data prior to creating the software option number.
1)
Ensure that you have the latest revision of application software.
2)
Follow the procedures in Appendix M for loading the application software.
3)
At the
HFA II - i
Main Menu, select the
“i”
button at the top of the screen. When the
HFA II - i
Unit Configuration screen is displayed, verify that the correct Model Number,
Hardware Option Number and Software Option Number have been set, based on
instrument model (see Table G-1).
NOTE:
The Hardware and Software Option Numbers are based on revision C3 software
residing on the hard disk.
Table G-1.
Model Number
Hardware Option
Number
Software Option Number
720
i
000134
0011590C
740
i
230134
00157918
745
i
330134
0015f918
750
i
b30334
0017F9F8
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...