Parts Removal/Replacement
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
3 - 24
52235C1105
P
The connector at the end of the CRT neck is not keyed, be careful to position it
correctly. The key-notch in the connector should be located in the gap between pins
1 and 7 on the tube.
Follow-up Checks/Adjustments:
See Table 3-1.
FIGURE 3-4. CRT / Touch Screen Removal