Troubleshooting
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
5 - 10
52235C1105
Q/A Test Points Menu:
• Current Point - Not a selection, but a status window. Provides information as to the
parameters of the currently selected test point.
• Auto Sequence - The
HFA II - i
will automatically advance through all test points. (To
interrupt and terminate, press the patient response button and then select
Done
.
• Manual Sequencing - Each time the patient button is pressed, the test point will advance.
• Open Shutter - Opens the shutter.
• Close Shutter - Closes the shutter.
• Flash Shutter - Opens then closes the shutter.
• Initial Point - Presents a ten-key selection pop-up window. Allows a specific test point to
be entered.
• Next Point - Advances the test point.
• Previous Point - Moves to the previous test point.
• Done - Returns to the Verification Menu.