![background image](http://html1.mh-extra.com/html/zeiss/humphrey-hfa-ii-i-series/humphrey-hfa-ii-i-series_field-service-manual_3332218085.webp)
Humphrey Field Analyzer II - i
Parts Removal/Replacement
Confidential and Proprietary
52235C1105
3 - 39
Field Service Guide
3.28 Vertical (Y-axis) Turret Belt/Motor Removal
Caution —
Be careful not to damage or dirty the bowl.
1)
Remove the Y-axis turret belt from the large pulley.
2)
Remove the Y-axis turret belt from the small pulley.
Motor Removal:
3)
Remove the two screws securing the Y-axis motor cable bracket.
4)
Disconnect the motor cable at the connector.
5)
Loosen the two setscrews on the motor pulley and remove it from the shaft.
6)
Remove the four screws securing the Y-axis motor to the turret assembly.
7)
Remove the Y-axis motor.
Replacement Notes:
Adjust the belt tension (Section 4).
Follow-up Checks/Adjustments:
See Table 3-1.