Humphrey Field Analyzer II - i
Appendix C
Confidential and Proprietary
52235C1105
C - 11
Field Service Guide
C.7
HFA II - i
To Third Party Programs Serial Data Transfer
Transfer of serial data from the
HFA II - i
to a third party program such as Peridata© is possible
only in the Serial II transfer mode.
Transmitting
HFA II - i
Tests to Third Party Programs
Ensure that the Serial II Transfer Disk is loaded by checking the destination window of the
Transfer Files selection on the Files Menu. If not, load the Serial II Transfer Disk as described
in Appendix section C.2.
1)
From the HFA II - i Main Menu, select
System Setup
.
2)
Select the
Save/Transmit
option.
3)
Select
RS-232 Options
.
4)
On the RS-232 Options screen, select the following settings:
A) Set Baud Rate to 9600
B) Parity to Even
C) Data Bits to 7
D) Stop Bits to 1
5)
Press
Proceed
. The Transmit/Save screen will then appear.
6)
Press the
Main Menu icon
in the upper right corner of the screen.
7)
From the Main Menu, select
File Functions
.
8)
Select
Transfer Tests
. A dialog box titled “Disk Options” will appear on screen.
9)
Select
Hard Drive
as the Source.
10)
Select
HFA I Serial Cable
as the Destination.
11)
Select the
Directory Order
(either
Name
or
Date
), then press
Proceed
.
• If you chose
Name
:
To select all files on the drive, press
Enter
, and
Select All
. All files on the
disk will be marked with a “
9
” to indicate that they have been selected.
• If you search files by patient name:
Enter the name - or just the first few characters - of the patient’s last name whose test(s)
you wish to transfer. Select the test(s) you wish to transfer. All files selected will be
marked with a “
9
” to indicate that they have been selected.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...