![background image](http://html1.mh-extra.com/html/zeiss/humphrey-hfa-ii-i-series/humphrey-hfa-ii-i-series_field-service-manual_3332218275.webp)
Humphrey Field Analyzer II - i
Appendix E
Confidential and Proprietary
52235C1105
E - 1
Field Service Guide
Appendix E. Instrument/BIOS Configuration
E.1 Instrument Configuration
This section provides a listing of configuration settings that are critical to the operation of
the
HFA II - i
unit.
Bootup Sequence
In the Advanced CMOS SETUP, the first boot device is IDE-0 (hard drive). The
instrument will not boot up a system disk from either the Magnetic Optical drive or the
floppy drive (if configured with the default settings). Special instructions have been
programmed in the instrument to recognize the installation disks and the Diagnostic
Support Tool disks. If an application diskette or the DST diskette are in the floppy drive
during instrument power up, the system will recognize them and allow the system to boot
from the floppy disk. If you attempt to load any other diagnostic software utilities that have
a system on the disk, you will need to first set the first boot device to floppy.
Note:
If you set the first boot device in the Advanced CMOS Setup to Floppy for
diagnostic purposes, it is necessary to reset the boot sequence to IDE-0 after using the
utility. Failure to do so will create a “Non-system disk or disk error Replace and strike any
key” any time a floppy disk is left in the floppy drive during the bootup sequence.
Changing the
BIOS Advanced CMOS Setup First Boot Device
1)
Set the
HFA II - i
system power switch to
ON
.
2)
Press and hold the
Delete
key.
3)
A
Enter current Password
dialogue will appear.
4)
Enter the password.
5)
The
AMIBIOS HIFLEX SETUP UTILITY
screen will appear.
6)
Use the down arrow key to select
Advanced CMOS Setup
.
7)
Press
Enter
.
8)
The
AMIOBIOS Setup - Advanced CMOS Setup
screen is displayed.
9)
Use the down arrow to select
1
st
Boot Device
.
10) Use the
PgUp
(page up) or
Pg Dn
(page down) keys to select the first boot device. Set the
first boot device to floppy if you plan on using the diagnostic utilities. Set the first boot
device to IDE-0 if you plan on resetting the instrument for normal operation.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...