Humphrey Field Analyzer II - i
Adjustment / Calibration
Confidential and Proprietary
52235C1105
4 - 31
Field Service Guide
NOTE:
Push the Tool
Stand with your finger as
shown or use the Fore-
Head Rest Clip Tool as
shown in the Appendix.
FIGURE 4-10. Proper Tool Stand Positioning
Pre-Setup (for all
HFA II - i
models)
1)
Push the tool stand forward slightly on the chinrest so that the metal tips on the Tool
Stand Alignment Adaptor make contact with the headrest (Figure 4-10). Then raise the
trial lens holder and pull it forward. This is the position that the tool stand should be in
when performing the procedure.
2)
The right-angle extension on the fake eye assembly will press against the trial lens
holder and position it correctly (Figure 4-11). (Set the chinrest height so that the right-
angle extension touches the trial lens holder at about the 0 - 0 level on the trial lens
holder scale.)
NOTE:
Once you have the trial lens holder set, do not move it.
3)
From the Calibration main menu, select
Camera
, then
Position
, and then
Camera
Monitor.
4)
By using the menus, move the trial lens holder image left or right, until the left outer
edge of the trial lens holder image is touching the left edge of the video insert; then
select
X-5
(5 more steps to the left), and
STORE
.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...