Humphrey Field Analyzer II - i
Appendix G
Confidential and Proprietary
52235C1105
G - 3
Field Service Guide
G.1 Initializing the Cal/Config Data
This procedure is required whenever a new hard drive has been installed, before calibration
is performed.
1)
Set up the DST for use as described in Appendix A.2, under Diagnostic Support Tool.
2)
From the DST Main Menu, select
Configuration
. The Configuration Menu
appears.
3)
Select
Initialize the Configuration
. A warning message appears informing
you that all configuration and calibration data will be destroyed.
4)
Select
OK
. A test log screen appears confirming whether or not you want to continue.
Select
Initialize
to erase all contents of the Cal/Config Data. (Select
Quit
to
return to the Configuration Menu.)
5)
“Initialization successful” message will appear.
NOTE:
If you initialize the CAL/Config Data, you must also perform the procedures for
Setting Serial Number
,
Setting the Model/Hardware Options
, and
Setting the Software
Options
described below. Then recalibrate the instrument (see Table 3-1).
G.2 Setting Serial Number
This procedure is required any time the hard drive is replaced or the Cal/Config Data is
initialized. During this procedure you will be instructed to type in the serial number of the
instrument. The serial number can always be found on the label located on the rear cover
assembly.
NOTE:
The instrument’s serial number NEVER changes throughout the life of the
instrument. Instrument serial numbers are critical for tracking instrument reliability,
performance, and customer satisfaction.
1)
Set up the DST for use as described in Appendix A.2, under Diagnostic Support Tool.
2)
From the DST Main Menu select
Configuration
. The Configuration Menu appears.
3)
Select
Set Serial Number
, then select
Edit
. A ten-key pad appears. Type in the
four-digit model of the instrument. For example, 0750, then select [Enter].
4)
A ten-key pad appears. Type in the six-digit serial number of the instrument. For
example, 001000, then select [Enter].
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...