Troubleshooting
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
5 - 20
52235C1105
System startup state 16
- Verify Wedges failed
Recommended Action:
• Dim room lights and retry
• Perform shutter and wedge calibration
• Check for broken, cracked, or very dirty green and/or blue filters in the detector assy.
replace as necessary
• Check/replace ND Wheel
System startup state 17
- Projector off failed
Recommended Action:
• Replace projection lamp and restart
• Check/replace projection lamp cable
• Check projection path alignment
• Check/replace spot detector/cable