Humphrey Field Analyzer II - i
Appendix A
Confidential and Proprietary
52235C1105
A - 11
Field Service Guide
A.2.19
How To Set Up The P-Factor Diskette
A
"Read P Factors"
button has been added to the
Light Intensity
menu. Selecting the
"Read P
Factors"
button from the
Light Intensity
menu causes the Soligor meter calibration P factor values
to be read from a disk (located in the
HFA II - i
floppy drive) and entered into the menu. The
Service Engineer must create the Soligor meter calibration P factors disk. The P Factors disk is
created by copying a file with the name
"pfactors.cfg"
on a DSHD floppy disk. The file will
contain the Soligor meter P calibration factors for the representative’s meter. A sample file is
shown below. Save the file as an ASCII text file. Lines beginning with
'#'
are comments and are
ignored by the program. You must use all capital letters and the delimiter between the meter and
the serial number, all of the factors, and all of the factor values must be a TAB. As an example for
the meter, you would need to type
METER
(TAB)
1935
, where
(TAB)
represents pressing the TAB
key.
WARNING:
You must update your P Factors disk whenever
your light meter is recalibrated or replaced. Label and Date
the P Factors disk.
#################################
# pfactors.cfg
# Configuration file for MBC module to help enter lightmeter
# P Factors and Meter serial number
METER 1935
P1 393.32
P2 36.282
P3 8.187
P3B 6.849
P3Y 8.218
P3F 7.812
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...