Humphrey Field Analyzer II - i
Adjustment / Calibration
Confidential and Proprietary
52235C1105
4 - 3
Field Service Guide
4.1
Introduction
This section contains the procedures for adjustment and calibration of the
HFA II - i
.
It is important that you use Table 3-1 in Section 3 to determine when
adjustments/calibrations are required following removal/replacement of various parts
and assemblies of the
HFA II - i
.
The information contained here reflects the latest revision level of software as of the
time of this writing. Always check your Field Service Bulletins for possible
subsequent changes to these procedures.
Calibration and Adjustment Flow Chart
The flowchart shown on the following page provides an overview of all required
instrument calibrations and adjustments. Additionally, each block contains the
section number in this manual where detailed procedures can be found for that
specific calibration or adjustment.
Calibration/Diagnostics Access, Diagnostics Menu,
and
Calibration Menu
selections are
all software options. That is, each of these blocks is available via a menu screen.
Whereas, the
Other
blocks are all mechanically related adjustments and are not
available via software.
NOTE:
The Other blocks, labeled Trial Lens Holder, Camera Focus, and Bowl IR
LEDS/Mirror, are mechanical adjustments that must be checked/performed prior to
performing the related software calibrations of Position and Intensity.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...