Humphrey Field Analyzer II - i
Appendix P
Confidential and Proprietary
52235C1105
P - 1
Field Service Guide
Appendix P. Cable Diagrams
The Data Transfer cables are frequently used with the
HFA II - i
. Shown below are the
cable diagrams for troubleshooting or fabrication purposes. The cables can also be
ordered directly from Carl Zeiss Meditec Parts Department.
Data Transfer Cables
HFA I to HFA II - i Serial Cable (P/N 52416)
- The HFA I to HFA II - i null
modem serial cable is used when transferring patient data files from an HFA I to an
HFA
II - i
. Refer to Appendix C for details on cable hookup and use. The cable is available
in a ten foot length.
HFA I End (25-Pin Male)
HFA II - i
End (9-Pin Male)
2 2
3 3
4 8
5 7
6 4
7 5
20 6