![background image](http://html1.mh-extra.com/html/zeiss/humphrey-hfa-ii-i-series/humphrey-hfa-ii-i-series_field-service-manual_3332218034.webp)
General Information
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
1 - 22
52235C1105
C
Uninterruptible Power Supply (UPS)
Carl Zeiss Meditec does not sell a UPS (provides battery backup during power loss). However,
when selecting a UPS for use with the
HFA II - i
, note the following specifications and
recommendations:
C
The UPS must have a minimum 450 VA rating.
C
It must be dedicated to the
HFA II - i
. Do not connect the power table or other
devices into the UPS.
C
Recommended model: APC smart UPS 620.
Power backup during power loss will depend on the type of UPS selected. Consult a local
supplier for details.
C
HP LaserJet Printer
Carl Zeiss Meditec does not sell HP LaserJet Printers. Refer to Appendix D, Approved Parallel
Printers for details on using the HP LaserJet.
1.9
Specifications
HFA II - i
Field Analyzer
Physical Characteristics
Dimensions . . . . . . . . . . . . . . . . . . 23" H x 22.25" W x 19.25" D
(58.4 x 56.5 x 48.9 cm)
Weight . . . . . . . . . . . . . . . . . . . . . 87.6 lbs (36.3 kg)
Recommended Room Size . . . . . . 106" x 64" (269.2 x 162.6 cm)
(Ventilation recommended)
Environmental
Operating Temp (ambient) . . . . . . 41 to 95
E
F
Humidity (ambient) . . . . . . . . . . . . 20 – 85%
Heat Generated . . . . . . . . . . . . . . 1,570 BTU/Hr (460 W)
Electrical
Input Power . . . . . . . . . . . . . . . . . 100 – 120 V @ 4 A, 50-60 Hz
230 V @ 2 A, 50 Hz
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...