Humphrey Field Analyzer II - i
Appendix M
Confidential and Proprietary
52235C1105
M - 3
Field Service Guide
21) When the instrument powers up to the main menu, select the ”
i
” button from the
top of the main menu. Verify that the correct revision of software has been loaded.
The software revision level can be viewed opposite the “Operating System” listing.
22) The installation is now completed.
NOTES:
If the system software was loaded on a newly initialized drive, you will
encounter a number of system errors and messages before the main menu is displayed.
One of the messages will indicate that a new patient database will be created.
If the software was loaded on a newly initialized drive you will need to restore the
calibration configuration, unit configuration, patient database and any software
options that were on the unit prior to the initialization. See section 4.8 for instructions
on restoring the calibration configuration.