PM and System Checkout
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
2 - 4
52235C1105
The hardware options number corresponds to the model number of the instrument and is
stored in the instrument's configuration file on the Hard Drive. If the hardware options
number is set as a model 750i, all software features are enabled. If set for a 740i, only those
features intended for the 740i are enabled. If set for a 720i, only those features intended for
the 720i are enabled.
The hardware options number should be verified during the system checkout procedure.
Refer to Appendix G for specific hardware options numbers.
The hardware options number also appears on the System Log printout and the Cal/Wedge
printout. You can print this information using the following procedures:
For the System Log
•
From the main menu, select
System Setup
; then select
Print System Log
. The
hardware options number and instrument serial number are printed on the line
titled
Machine ID
, at the beginning of the printout.
For the Cal/Wedge printout
•
From the Calibration main menu, select
Print Cal Values
to print out the instrument
calibration data. The hardware options number and instrument serial number are
printed on the line titled
Machine ID
, near the beginning of the printout.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...