3.26 Horizontal (X-axis) Turret Motor Removal
1)
Remove the front and rear covers (3.2, 3.3).
2)
Disconnect the "X-Axis" 6-pin connector (P10) from the Projection Interconnect
PCB.
3)
Remove the projection assembly and install it on the projection mount tool (3.17).
4)
Remove the four screws securing the X-axis turret motor to the projection
assembly.
5)
Remove the belt from the motor pulley; then remove the motor.
6)
Loosen the two setscrews on the hub/pulley and remove it from the motor shaft.
Replacement Notes:
Adjust the belt tension (Section 4).
Follow-up Checks/Adjustments:
See Table 3-1.