5160 Hacienda Drive, Dublin, CA 94568 (925) 557-4100 (1-800) 858-1508
Date ________________________
Dear Doctor:
We checked the calibration of your Field Analyzer, serial number ____________, today
and found that it was reading dB too high
0
/ low
0
(select). The
instrument has now been recalibrated.
Please keep this calibration shift in mind when comparing future results against
previous field. If it was reading too high, future tests of stable patients will tend to show
an artifactual loss in threshold sensitivity in the amount shown above. If it was reading
too low, future tests will tend to show an artifactual improvement.
We recommend yearly preventative maintenance and calibration checks for your Field
Analyzer. Inquire about our service contracts, which include a yearly preventative
maintenance visit.
If you have any questions, please contact Carl Zeiss Meditec at 1-877-486-7473 and
ask to have your area’s Regional Service Manager give you a call.
Sincerely,
Carl Zeiss Meditec
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...