Page 4
FA2i-045A
4) Insert the 5-prong plug into the power receptacle on the lift mechanism as shown in Figure 5.
Figure 5 Insert the 5-Prong Plug
Notes:
Prior to installing the HFA table top onto the support tray:
•
Ensure that the correct fuses are installed for the power mechanism to be tested.
•
Plug the power cord into the power inlet on the lift mechanism and then plug the power cord into
any available power source (see label on the power lift mechanism for the proper power input
requirements).
WARNING –
Do not touch any exposed wire or terminal while the power cord is
plugged into the power source.
Notes:
•
Press the up and down button to test the lift mechanism to ensure that the 5-prong plug is
properly seated in the receptacle.
•
During testing, press the up button and allow the lift mechanism to reach its maximum extension.
Disconnect the power cord from the power inlet and observe that there is no slippage of the lift
mechanism.
•
Steps 5 and 6 require two people to perform the task.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...