TROUBLESHOOTING
Normally rebooting the printer and the HFA2i will restore printing operation. If rebooting does not
restore normal printing, run the self test function. The self test function will often restore normal
printing.
The self test may be run with or without the data cable connected to the HFA2i. To initiate the self
test feature:
1)
Set the printer ON/OFF switch to OFF.
2)
Press in and hold the paper advance button.
3)
Set the printer ON/OFF switch to ON.
4)
Release the paper advance button.
5)
The printer will produce a printout that includes:
a) Model information
b) Firmware version
c) Head test
d) Platen test
e) Character sets
f) Buffer information
g) Analog input
Note –
Be sure to attempt the self test and retry printing prior to replacing a printer.