3.24 Shutter Motor Removal
1)
Remove the projection assembly and install it on the projection mount tool (3.17).
2)
Disconnect the shutter cable (P6) from the Projector Interconnect PCB (cut the
cable ties if necessary).
3)
Remove the two screws securing the shutter motor to the projection assembly,
and remove the motor.
4)
Loosen the setscrew securing the hub and shutter to the motor shaft, and remove
them from the shaft.
Replacement Notes:
P
Position the shutter for adequate clearance between the glass wedge and the
projection casting. Gently bend the shutter, if necessary.
Follow-up Checks/Adjustments:
See Table 3-1.
3.25 Carriage Motor Belt Removal
1)
Loosen
the four screws securing the carriage motor to the projection assembly.
2)
Remove the two small screws securing the belt plate against the belt.
3)
Remove the belt.
Replacement Notes:
Adjust the belt tension (Section 4).
Follow-up Checks/Adjustments:
See Table 3-1.
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...