Parts Removal/Replacement
Humphrey Field Analyzer II - i
Confidential and Proprietary
Field Service Guide
3 - 14
52235C1105
FIGURE 3-2. Operator Panel Bottom Mounting Screws
Replacement Notes:
P
Be sure to slide the ribbon cables back into their clips.
P
Make sure that the patient response button connector module is mounted in the
bottom pan,
with its contacts positioned upward
, before you reinstall the
operator panel assembly.
Follow-up Checks/Adjustments:
See Table 3-1.