FA2i-012
Page 11
FIGURE 6. Chinrest and Forehead Rest Alignment Check
S e t t i n g t h e H a r d w a r e O p t i o n s N u m b e r a n d t h e S o f t w a r e O p t i o n s N u m b e r
S e t t i n g t h e H a r d w a r e O p t i o n s N u m b e r a n d t h e S o f t w a r e O p t i o n s N u m b e r
1)
Use the Diagnostic Support Tool (DST) diskette to set the Hardware Options number:
From the DST Main Menu select
Set Model/Hardware Options
. Select
Set Model
745i Defaults
, a test log menu appears and displays the Model 745i default settings.
Select
Continue.
2)
Load the current revision of HFA II-i System software included in the upgrade kit (refer to the
Appendix section of the Service manual). Loading the software will set the software options
number that is stored on the Hard Drive. This step must be done
AFTER
the hardware options
number has been set.
3)
Once the software is loaded (see Note below), proceed to the Setup menu and print out the
system log. Verify that the correct hardware options number has been entered. (The
hardware options number is displayed at the top of the printout, adjacent to the machine ID.)
NOTE:
The instrument will probably give an error during the power-up sequence. To bypass the error, select
Proceed to get to the Main Menu. At the Main Menu, the Setup menu option can be selected. [When new hardware is
installed (color wheel, yellow lamp assembly, etc.), an error will occur during the power-up sequence because the new
hardware is not yet calibrated. This calibration will be performed in the next procedure.]
4)
Insert the configuration backup diskette in the floppy drive and select the
Restore
Configuration
pad in
Additional Setup
. (Once complete, the printer selection and Alter
Summary of Contents for humphrey HFA II-i series
Page 4: ......
Page 179: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 424: ...Page 4 FA2i 013A PROPRIETARY and CONFIDENTIAL...
Page 502: ...FA2i 041 Page 3 Figure 2 Press rectangle in Top Left corner...
Page 516: ...Page 2 FA2i 046 Figure 1 Glass Film Wedge Detectors P N 67417 Note Spacing Washer Placement...
Page 517: ...FA2i 046 Page 3 Figure 2 Y Motor Turret Detector P N 67418 Note Spacing Washer Placement...