Chapter 15: Maintaining Your Library
Maintenance Actions
Quantum Scalar i6000 User’s Guide
635
Horizontal Alignments
The Horizontal Alignments graphical report shows the results for tests
of horizontal alignment of tape magazines on the drive-side and door-
side across frames, and for horizontal alignment of I/E stations across
frames.
Note:
This graphical report is not generated for libraries with only
one frame.
• If the graphical report shows a failed result for the drive-side or
door-side, make sure that all tape magazines are installed properly
on that side and that the calibration targets are correctly snapped
on to the magazines.
• If the graphical report shows a failed result for the I/E station, make
sure the I/E station and front door are completely shut.
• If running the test again still generates failed results, realign the
middle X-axis rail and check the alignment of the top and bottom X-
axis rails at the location of the failure.