C h a p t e r 9 P R O C E S S R E C I P E
7/26/00
INTEGRA Operations Revision A
9-49
20-0200-231 (CR), 20-0200-232 (Std), 20-0200-251 (CD)
Change DOSE CHEM
47 Toggle the DOSE CHEM ON or OFF (default: OFF) for Brush
Station #1 or Brush Station #2.
When DOSE CHEM is set to ON, process chemical is used to dose
the brushes.
Brush dosing continues for the time set in the WAFER
DOSE TIME.
When DOSE MODE is selected, and/or the DOSE LOW FLOW
DI, and/or the DOSE HI FLOW DI are set to ON, then they will
both start when the dosing chemical starts.
Change DOSE LOW FLOW DI
48 Toggle the DOSE LOW FLOW DI ON or OFF (default: ON) for
Brush Station #1 or Brush Station #2.
When DOSE LOW FLOW DI is set to ON, high flow DI water is
used to dose the brushes.
Brush dosing continues for the time set in
the WAFER DOSE TIME.
Only systems configured with Low Flow DI water air control
valves use this parameter. Systems without the Low Flow DI
water air control valves do not use this parameter and will
continue to dispense DI water to the brushes at the DI Low Flow
meter setting.
When DOSE MODE is selected, and/or the DOSE LOW FLOW
DI, and/or the DOSE HI FLOW DI are set to ON, then they will
both start when the dosing chemical starts.
STANDARD
Summary of Contents for Synergy Integra
Page 8: ......
Page 24: ...1 4 INTEGRA Operations Revision A 7 26 00 20 0200 231 CR 20 0200 232 Std 20 0200 251 CD...
Page 38: ...2 14 INTEGRA Operations Revision A 7 26 00 20 0200 231 CR 20 0200 232 Std 20 0200 251 CD...
Page 222: ...9 62 INTEGRA Operations Revision A 7 26 00 20 0200 231 CR 20 0200 232 Std 20 0200 251 CD...
Page 314: ...10 76 INTEGRA Operations Revision A 7 26 00 20 0200 231 CR 20 0200 232 Std 20 0200 251 CD...