System Operation
3-10
Note:
The
Visual
Instrument Setup
will vary, depend-
ing on the in-
strument configu-
ration.
5
Open the
Visual Instrument Setup
dialog box:
6
Click on the
grating for the
excitation
monochromator.
The
Grating/Turret
dialog box appears.
7
Click on the
Calibrate...
button.
This opens the
Enter Correct Position
dialog box:
8
Enter the actual xenon-
lamp peak, 467 nm.
9
Click OK.
The excitation monochromator should
now be calibrated.
10
Click Close to close the
Grating/Turret
dialog
box.
11
Close
Visual Instrument Setup
.