Fluorolog-3 v. 2.2 (31 Jul 2002)
Troubleshooting
7-7
Stray light
In the following diagram, notice the high level of stray-light below 380 nm in the water
Raman spectrum.
To correct this problem,
1
Inspect the cuvette surface for fingerprints and
scratches.
2
Clean the cuvette or use a new one.
3
Verify that the excitation and emission slits are
set correctly for a water Raman scan.
4
Verify that the excitation spectrometer is at the
correct position.
Nor
mal
S
High stray light in a water Raman scan.
0
10
20
30
40
50
60
365
380
395
410
425
440
455
Wavelength (nm)