Fluorolog-3 v. 2.2 (11 Jul 2002)
System Operation
3-1
Note:
The boot disk should access
for ~ 60 s to load system drivers.
Warning:
When the lamp is
turned on, a large voltage is put
across the lamp, during which a
spike can feed back down the
electrical line. This spike can
cause damage to computer
equipment if the equipment is
operating at the time and on the
same power circuit as the
Fluorolog
®
-3, when the lamp is
started.
3: System Operation
Turning on the system
1
Start the lamp.
The lamp must be turned on
prior to the Fluorolog
®
-3,
accessories, and peripheral
equipment.
a
On the back of the lamp
housing, turn on the switch
marked “power.”
b
Just above the “power” switch, turn on the “main lamp” switch.
2
Start the fluorometer accessories.
Turn on any automated accessories (e.g., Temperature Bath, MicroMax,
etc.) used with the Fluorolog
®
-3.
3
Start the SpectrAcq.
a
Make sure the boot disk is in the floppy
drive.
b
Push in the power button to start.
Immediately below the power button, the
green LED indicator lamp should turn on.